• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CORDIS US. CORP SABERX014; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS US. CORP SABERX014; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number N/A
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/12/2025
Type of Reportable Event Malfunction
Event or Problem Description
As reported, the balloon of a saberx.014 pta, 2.5 x 200, 150cm, ruptured before reaching nominal pressure.There was no reported patient injury.The device was used in a below-knee (bk) case.The device will not be returned for analysis.
 
Additional Manufacturer Narrative
Due to system limitations, section h6 required to input type of investigation, investigation findings, and investigation conclusions for an initial report.Pending additional information to complete investigation.Additional information is pending and will be submitted within 30 days upon receipt.
 
Additional Manufacturer Narrative
After further review of additional information received the following sections have been updated accordingly: g3, g6, h1, h2, h3 and h6.As reported, the balloon of a saberx.014 pta, 2.5 x 200, 150cm, ruptured before reaching nominal pressure.There was no reported patient injury.The device was used in a below-knee (bk) case.Additional information was requested but not provided.The product was not returned for analysis.A product history record (phr) review of lot 2508020990 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported event of ¿balloon- burst at/below rbp¿ could not be verified, as the device was not returned for analysis.Consequently, the precise root cause cannot be determined.While vessel morphology and procedural factors, are likely contributors though not specified.Without the device for evaluation, a definitive link between the device and the reported event cannot be established.According to the instructions for use, which are not intended to mitigate risk, ¿carefully inspect the catheter prior to use to verify that the catheter has not been damaged and that its size, shape and condition are suitable for the procedure for which it is to be used.Flush or rinse all products entering the vascular system with sterile isotonic saline or a similar solution via the guide-wire access port prior to use.Never attempt to move the guide wire when the balloon is inflated.Do not advance the catheter against significant resistance.The cause of resistance should be determined via fluoroscopy.Inflation in excess of the rated burst pressure may cause the balloon to rupture.Use the recommended balloon inflation medium (25% contrast medium/75% sterile saline solution).It has been shown that a 25/75% contrast / saline ratio has yielded faster balloon inflation / deflation times.Never use air or other gaseous medium to inflate the balloon.If resistance is felt during post procedure withdrawal of the catheter through the introducer sheath, determine if contrast is trapped in the balloon with fluoroscopy.If contrast is present, push the balloon out of the sheath and then completely evacuate the contrast before proceeding to withdraw the balloon.If resistance is still felt during post procedure withdrawal of the catheter, it is recommended to remove the balloon catheter and guidewire/ introducer sheath as a single unit.Do not continue to use the balloon catheter if the shaft has been bent or kinked.¿ neither the phr nor the information available suggests a design or manufacturing related cause for the reported event.Therefore, no corrective or preventive action will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SABERX014
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
CORDIS US. CORP
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer (Section G)
CORDIS US CORP.
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer Contact
karla castro
15th building smes venture park songshan lake hi-tech zone
dongguan, guangdong 
7863138372
MDR Report Key24008125
Report Number1016427-2026-00002
Device Sequence Number4810241
Product Code LIT
UDI-Device Identifier20705032089347
UDI-Public(01)20705032089347(17)280731(10)2508020990
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K221832
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number47525200M
Device Lot Number2508020990
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/17/2025
Supplement Date Received by Manufacturer01/16/2026
Initial Report FDA Received Date01/08/2026
Supplement Report FDA Received Date01/23/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/02/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GLADIUS.; PARENT SELECT 5FR.
Patient SexUnknown
-
-