|
This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional narrative: h3, h6: part: 03.043.028 lot: 21413102 manufacturing site: werk selzach supplier:(b)(4) release to warehouse date: 01 september 2021 expiration date: na a manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.The product was returned to depuy synthes for evaluation.Visual inspection of the returned device found the weld of the pin is broken, pin was returned detached from de driving cap, distal sleeve was loose as well.A complete functional test cannot be performed due to breakage and detached pin, however, due to this condition the spring and pull button were loose confirming the complaint condition.Note: a driving cap from family 03.043.028 was evaluated and investigated by r&d.The observed condition of the pin has been previously assessed by materials and testing.Materials and testing provided drafted report that showed some deficiencies in the weld.Conclusion from r&d was that although the weld might not be perfect there must be some significant load leading to the failure of the weld.Materials and testing subsequently performed a finite element analysis to investigate origin of stress at the interface between cross pin and pull button.Analysis showed that strong off axis hammer blows can lead to oscillations in the system that can cause stress that can lead to breakage of the laser weld at the end of the pin and subsequent migration.Surgical technique guide mentions at page 16: to use the hammer, attach the driving cap to the insertion handle and secure it by twisting it one quarter turn.Apply light and controlled hammer blows to seat the nail.Notes: the hammer guide may aid in controlling the direction of the hammer blows.Therefore, the hammer guide can be attached to the back end of the drive cap by screwing both parts together.A dimensional inspection was unable to be performed due to the post-manufacturing damage.The overall complaint was confirmed as the observed condition of the driving cap would have contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.The drawings reflecting the current and manufactured revisions were reviewed.
|