| Catalog Number |
PMW35 |
| Medical Device Problem Code |
Mechanics Altered (2984)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/15/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during the unk surgery, after fired, noted the staples malformed.Changed the new one to complete surgery.There was no patient consequence reported.No additional information can be provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 1/8/2026.D4: batch # unk.The device/photo upon which this medwatch is based has been received, however, the evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformance were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 2/6/2026.Investigation summary.The product was returned to ethicon endo-surgery for thorough evaluation, which included visual inspections of the returned device.Visual analysis of the returned sample revealed that the pmw35 device was returned nonfunctional.The device was received with the cartridge partially detached from the handle due to insufficient handle weld on the upper handle directors.No functional test was performed due to the condition of the device, however the insufficient handle weld could have cause firing issues when the cartridge began to detached from the handle.The separation of device has been correlated to the manufacturing process.Based on the information currently available, a product issue was identified during the investigation of the sample received.This product issue will be addressed through ethicon endo surgery¿s quality system.Device history review.A manufacturing record evaluation was performed for the finished device lot 669d43, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|