Additional information: d9, g3, h3, h6.The complaint allegation was confirmed.One unpackaged ar-3641-1 2.6 mm knotless fibertak¿ anchor, batch 15424553, was received for evaluation.Visual evaluation of the inserter identified no damage.Evaluation of the anchor identified fraying and deformation of the sheath/anchor.The most likely cause of the reported failure is use error, including improper bone preparation and incorrect surgical technique with the device.Directions for use dfu-0054 at revision 8 bio-fastak, fastak¿, suturetak®, and fibertak® suture anchors.E.Warnings: 7.Avoid excessive impaction during anchor insertion as this could lead to inserter damage and/or breakage.If insertion resistance is encountered, do not impact harder.Replace the implant and repeat the drilling/insertion process.G.Precautions: 2.Surgeons are advised to review the product-specific surgical technique prior to performing any surgery.Arthrex provides detailed surgical techniques in print, video, and electronic formats.The arthrex website also provides detailed surgical technique information and demonstrations.Or contact your arthrex representative for an onsite demonstration.4.Fastak and fibertak suture anchors only: it is important to stop drilling as soon as the chuck contacts the guide or grasper.Failure to do so may result in damage to the suture and/or implant.5.Insert the anchor with the same orientation as that of the prepared drilled bone hole to avoid damage to the anchor or inserter.
|