Model Number 2.6 MM KNOTLESS FIBERTAK ANCHOR
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/17/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
It was reported that during an arthroscopic surgery the device didn't work.Per complaint reporter there was no harm for patient, operator or third party.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.Update 30-dec-2025: it was further reported that the anchor did not deploy correctly and the device pulled out of the prepared bone channel.
 
Additional Manufacturer Narrative
Additional information: d9, g3, h3, h6.The complaint allegation was confirmed.One unpackaged ar-3641-1 2.6 mm knotless fibertak¿ anchor, batch 15424553, was received for evaluation.Visual evaluation of the inserter identified no damage.Evaluation of the anchor identified fraying and deformation of the sheath/anchor.The most likely cause of the reported failure is use error, including improper bone preparation and incorrect surgical technique with the device.Directions for use dfu-0054 at revision 8 bio-fastak, fastak¿, suturetak®, and fibertak® suture anchors.E.Warnings: 7.Avoid excessive impaction during anchor insertion as this could lead to inserter damage and/or breakage.If insertion resistance is encountered, do not impact harder.Replace the implant and repeat the drilling/insertion process.G.Precautions: 2.Surgeons are advised to review the product-specific surgical technique prior to performing any surgery.Arthrex provides detailed surgical techniques in print, video, and electronic formats.The arthrex website also provides detailed surgical technique information and demonstrations.Or contact your arthrex representative for an onsite demonstration.4.Fastak and fibertak suture anchors only: it is important to stop drilling as soon as the chuck contacts the guide or grasper.Failure to do so may result in damage to the suture and/or implant.5.Insert the anchor with the same orientation as that of the prepared drilled bone hole to avoid damage to the anchor or inserter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.6 MM KNOTLESS FIBERTAK ANCHOR
Common Device Name
NON-DEGRADABLE FIXATN FASTENER
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key24011139
Report Number1220246-2026-00107
Device Sequence Number9250645
Product Code MBI
UDI-Device Identifier00888867290952
UDI-Public(01)00888867290952(11)250422(17)300331(10)15424553
Combination Product (Y/N)N
Initial Reporter CountryPL
PMA/510(K) Number
K251809
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 04/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2.6 MM KNOTLESS FIBERTAK ANCHOR
Device Catalogue NumberAR-3641-1
Device Lot Number15424553
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/19/2025
Supplement Date Received by Manufacturer04/23/2026
Initial Report FDA Received Date01/08/2026
Supplement Report FDA Received Date04/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/22/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-