RESPIRONICS, INC. REMSTAR AUTO A-FLEX; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number AH560HS
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/31/2025
Type of Reportable Event Malfunction
Event or Problem Description
A remstar auto a-flex device was returned to a third-party service center as part of the uno recall process with no initial alleged complaint.During the evaluation of the device, the service technician observed visible foam particles inside the blower kit and object code indicates the blower contributed to the foam degradation.Additionally, the service technician noted dust contamination and error codes were displayed (e020 | e-20: err_motor_spinup_flux; e031 | e-31:err_motor_vbus_high_blower_off; e053 | e-53: err_comp_log_sem_timeout) in the error log.The device was scrapped by the service center after the evaluation was completed.This event is reportable under fda 21 cfr part 803 as it meets the criteria for a serious injury and/or product malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REMSTAR AUTO A-FLEX
Common Device Name
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key24012082
Report Number2518422-2026-001256
Device Sequence Number9343677
Product Code BZD
UDI-Device Identifier00606959005860
UDI-Public(01)00606959005860
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K131982
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 01/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberAH560HS
Device Catalogue NumberAH560HS
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/31/2025
Initial Report FDA Received Date01/08/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/27/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1974-2021
Patient Sequence Number1
Patient SexUnknown
-
-