| Model Number |
VTICM5_ 12.6 (-16.50/ 1.5) |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Blurred Vision (2137)
|
| Date of Event |
11/09/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H11- manufacturer narrative: each lens is subjected to a 100% assessment of the power (spherical and cylinder) and optical resolution during the manufacturing process in order to determine acceptability per the lens model and diopter.Secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.H6- investigation type code: 4110- lens work order search-no similar complaint event(s) within associated lots were found.Claim# (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced a refractive surprise.In a separate surgery the lens was exchanged with the same model/ length lens, different power and the problem was resolved.Reportedly, "s-mr and cr has big different number".The cause of the event was reported as unknown.There are multiple medical device reports associated with this patient.This report is 1 of 2 total reports.Based on the information provided, the risk assessment for our product, and our experience, we have determined the cause or contribution to the reported issue is not indicative of a manufacturing related issue or product deficiency.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: h3: type of investigation code: 10- device evaluation: the lens was returned dry in a microcentrifuge vial with residue/debris on product.Visual inspection found no visible damage with fibre on lens surface.Dimensional inspection found the lens to be within specifications.Functional inspection found the returned lens did not meet original values measured at the time of manufacturing.Claim #: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: h6- type of investigation code: 3331- device history record (dhr) review: based on the results of the investigation, all released devices from the associated work order(s), including the suspected device, have been manufactured within established process parameters; and there is no indication that the manufacturing and processing of the device contributed to the complaint issue.Claim# (b)(4).
|
| |
|