| Model Number |
EGIA60CTAVM |
| Medical Device Problem Codes |
Retraction Problem (1536); Noise, Audible (3273)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
09/08/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during the procedure, the reload made a strange noise.The issue was resolved by replacing the device and loading a new additional reload.Medtronic's initial evaluation of the incident device found that the i-beam was slightly advanced.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the reload had a full complement of staples.The i beam was slightly advanced.The proximal end of the reload adapter was broken.Functional testing was precluded due to the noted damage.It was reported that the reload made a strange noise.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue can occur due to over flexure of the reload while attached to the instrument.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|