| Model Number |
G150 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
| Date of Event |
12/11/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) was part of the system revision due to infection.Reportedly, the crt-d was used as a temporary pacing system.The patient was treated with intravenous antibiotics.No adverse patient effects were reported.The crt-d was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our post market quality assurance laboratory, additional information from product investigation indicates that the allegation against the device was not confirmed.The device was analyzed, passed all the baseline tests and exhibited normal device functions.This supplemental report is being filed to correct the d9: device avail for eval; d9: device return date; g2: report source; h2: follow-up type; h3: device eval by manufacturer; and h11: additional mfr narrative.
|
| |
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) was part of the system revision due to infection.Reportedly, the crt-d was used as a temporary pacing system.The patient was treated with intravenous antibiotics.No adverse patient effects were reported.The crt-d was explanted.
|
| |
|