|
Additional information provided in d.9., h.3., h.6.And h.11.No technical inquiries were received from the customer.One opened 0.9-millimeter phacoemulsification tip (phaco tip) loose in a blister was received for the report.The sample was visually inspected and found to be nonconforming with phaco tip bent at the middle area of the cannula and bent at the bevel with gouge marks going up the cannula and stop a little below the aspiration bypass (ab) hole.Wear was observed on threads, back of flange, or nut corners and consistent with threading on handpiece.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The investigation conducted a non-conformance review of the reported lot number.No deviations were identified and all corresponding production release specifications defined in the device master record were met.The returned sample was found to be visually nonconforming with a bent phaco tip.How and when the phaco tip became bent before the surgery could not be determined.Based on the evaluation of the information and materials received, the investigation was unable to identify the root cause or origin of the reported event.As the root cause and its origin are inconclusive, further investigative or manufacturing actions are not warranted at this time.However, an investigation has been initiated to determine root cause of the bent phaco tip.All phaco tips are 100% visually inspected by trained operators using 30x magnification during the manufacturing process.Any nonconformance is removed from the lot and scrapped.Complaint data for all company products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
|