| Catalog Number |
8065921501 |
| Medical Device Problem Code |
Dull, Blunt (2407)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/16/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A physician reported that an ophthalmic cutting knife was found to be blunt before the surgery.The details of the procedure were not provided.The surgery was completed on the same day without any patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.A sample is available that has not yet been received at manufacturing for evaluation the manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
One opened ophthalmic knife was received for the report of surgical blade is not sharp.The returned samples were visually inspected and was found to be conforming.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The investigation conducted a non-conformance review of the reported lot number.No deviations were identified and all corresponding production release specifications defined in the device master record were met.Based on the evaluation of the information and materials received, the investigation concluded that the root cause of the reported event is not determined as the returned sample was conforming for all visual testing.No further investigative or manufacturing actions are warranted at this time due to that the returned sample met specifications.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all manufacturer products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|