MEDOS INTERNATIONAL SÃ RL HD LAPSCP,EP,5.5,0,300; LAPAROSCOPE, GENERAL & PLASTIC SURGERY
|
Back to Search Results |
|
| Catalog Number |
242632 |
| Medical Device Problem Codes |
Display or Visual Feedback Problem (1184); Scratched Material (3020)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11: additional narrative: investigation summary = >¿ the product was not returned to depuy synthes, however a photo was provided for evaluation.Visual inspection of the returned photo found no observations pertaining to the nature of the reported event or any other anomaly.The overall complaint was not confirmed as the observed condition of the hd lapscp,ep,5.5,0,300 would have not contributed to the complained device issue.¿ based on the investigation findings, a potential cause cannot be established.Multiple factors are associated with this type of failure, and the complaint device needs to be physically received and evaluated in our service center to determine a potential cause of why the customer experienced the failure.It has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
|
| |
|
Event or Problem Description
|
|
It was reported that during an unspecified surgical procedure the hd lapscp,ep,5.5,0,300 device was found to have a large scratch and its image appeared blurry.It was not reported if there were any delays to the surgical procedure.It was reported that a spare device was used to complete the surgery.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11: additional narrative: h4: the device manufacture date has been added.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11: additional narrative: investigation summary the complaint device was received at the service center and evaluated.During the service evaluation the following defects were identified: visual : scratched/nicked, broken (2+ pieces) : chipped/shaved/delaminated, visual : deformed/bent and labeling : illegible etch.Per service report, this complaint was confirmed.Based on the investigation findings, the potential root cause is traced to user error.It has been determined that no corrective and/or preventative action is proposed and there is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review a manufacturing record evaluation was performed for the finished device serial number (b)(6), and no non-conformance was identified.
|
| |
|