• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH PERCUTANEOUS ARM 255MM; FORCEPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH PERCUTANEOUS ARM 255MM; FORCEPS Back to Search Results
Catalog Number 398.754
Medical Device Problem Codes Device-Device Incompatibility (2919); Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/19/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that when validating the equipment before sending it to the institution and when mounting the collinear gripper, it was observed that the hook presents resistance when mounted; once mounted, it is not possible to separate it from the clamp.
 
Additional Manufacturer Narrative
Product complaint # (b)(4) this report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: corrected: g1.H6 component codes: most relevant component code is g07002 (appropriate term/code not available) to capture no findings available due to no product returned.H6: the product was not returned to depuy synthes, however photos were provided for review.The photo investigation revealed that percut-arm f/collin reduc clamp , the provided evidence was not sufficient to confirm the reported event of unable to assemble/disassemble (2+ devices) & jammed/seized.Functionality issues cannot be evaluated through a photo investigation.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was not confirmed as the observed condition of the percut-arm f/collin reduc clamp would not contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot.Part: 398.754.Lot:180137-201.Manufacturing site: werk selzach.Supplier: (b)(4).Release to warehouse date: 21 sep 2018.Expiration date; na.A manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERCUTANEOUS ARM 255MM
Common Device Name
FORCEPS
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES SELZACH
bohackerweg 5
selzach 4434
SZ   4434
Manufacturer Contact
elba bello
1302 wright lane east
west chester, PA 19380
9083863534
MDR Report Key24019760
Report Number8030965-2026-00289
Device Sequence Number18780204
Product Code HTD
UDI-Device Identifier07611819283015
UDI-Public(01)07611819283015(10)180137-201
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number398.754
Device Lot Number180137-201
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/19/2025
Supplement Date Received by Manufacturer01/15/2026
Initial Report FDA Received Date01/09/2026
Supplement Report FDA Received Date01/19/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/21/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-