• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER-ENDOSCOPY PHOENIX USA; CATHETER, RECORDING, ELECTRODE, REPROCESSED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER-ENDOSCOPY PHOENIX USA; CATHETER, RECORDING, ELECTRODE, REPROCESSED Back to Search Results
Model Number 81104
Medical Device Problem Codes Mechanical Problem (1384); Use of Device Problem (1670)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/01/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to stryker sustainability solutions for evaluation.Upon inspection of the received complaint device, no relevant visible damage was detected.The catheter handle, deflection/ locking mechanism, shaft, connector, and distal tip appeared to be intact.Inspection of the handle revealed the deflection mechanism was very hard to manipulate due to the locking collar being excessively tightened.Loosening of the locking mechanism reduced some of the tension to the deflection, the device was not able to be manipulated effortlessly unless significantly loosening the locking collar.The device was evaluated for deflection.The catheter did not form a curve consistent with the visual work reference (vwr) specification.It is probable that the mechanical issue likely contributed to the reported event of, ".Cable didn't work when they plugged it in.Bad reprocessed stryker", as damage to the device's structural integrity (internal wiring) may adversely impact device performance.The inspection results determined that the complaint is confirmed for the reported failure mode.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.The reported event could be attributed to a mechanical issue as a result of mishandling subsequent to distribution, including shipping and storage conditions or improper manipulation of the device.The instructions for use (ifu) state: - do not attempt to use the reprocessed ep catheter prior to completely reading and understanding the directions for use.- do not alter this device.- inspect the packaging and catheter for damage or defects prior to use.- avoid excessive torque, stretching, kinking and/or bending of catheter, as this may interfere with distal tip shaping or cause damage to internal electrode wires.- avoid manual pre-bending of distal curve, as this may damage steering mechanism of steerable catheters.Handle catheter with care to avoid improper electrical functioning.- avoid excessive contact of handpiece with fluids, as this could adversely affect the electrical performance of the catheter.The reported event will continue to be monitored through post-market surveillance.
 
Event or Problem Description
It was reported that the device was bad.There was no patient injury or medical intervention and extended procedure time was reported however the complainant is not aware of the duration.These are commonly used devices that are readily available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Common Device Name
CATHETER, RECORDING, ELECTRODE, REPROCESSED
Manufacturer (Section D)
STRYKER-ENDOSCOPY PHOENIX USA
10232 s. 51st st.
phoenix AZ 85044
Manufacturer (Section G)
STRYKER-ENDOSCOPY PHOENIX USA
10232 s. 51st st.
phoenix AZ 85044
Manufacturer Contact
arsalan rizvi
1810 w. drake drive
tempe, AZ 85283
8888883433
MDR Report Key24019812
Report Number0002090040-2026-00002
Device Sequence Number15691400
Product Code NLH
UDI-Device Identifier00885825009557
UDI-Public00885825009557
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K112232
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 01/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number81104
Device Catalogue Number81104
Device Lot Number5533256
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/31/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/19/2025
Initial Report FDA Received Date01/09/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/17/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-