• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. DJO SURGICAL; TAPERFILL HIP STEM, LATERAL OFFSET, SIZE 8

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. DJO SURGICAL; TAPERFILL HIP STEM, LATERAL OFFSET, SIZE 8 Back to Search Results
Catalog Number 425-97-008
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Pain (1994); Periprosthetic fracture (4902)
Date of Event 10/07/2025
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery due to post-operation pain and periprosthetic fracture.
 
Additional Manufacturer Narrative
See d4 expiration date, d10, h4, h6 and h11.The agent reported "(the patient complained of pain since day three post operation.She was found to have a peri-prosthetic fracture of her femur.)".The previous surgery and the surgery detailed in this event occurred 14 days apart.This evaluation is limited in scope as the items associated with this investigation were not returned to djo surgical - austin for examination.The surgery was completed as intended and without incident.If additional information regarding the reported event is submitted at a future date, this investigation will be re-evaluated.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.There were no findings during this evaluation that indicate the reported devices were defective.The surgeon performed this procedure to remedy the patient's condition.No further action is deemed necessary.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.The revision surgery was completed successfully.The root cause of this complaint was a revision surgery due to pain and peri-prosthetic fracture.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.Due to short time between previous and revision surgery, it is possible that the event may have occurred due to lack of post-operative care, patient noncompliance with medical instructions or incorrect implant selection, patient activities or trauma.There are multiple factors that may also contribute to an event that are outside the control of djo surgical.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DJO SURGICAL
Common Device Name
TAPERFILL HIP STEM, LATERAL OFFSET, SIZE 8
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758
Manufacturer Contact
stacey bonnell
9800 metric blvd
austin, TX 78758
5128346313
MDR Report Key24024451
Report Number1644408-2025-01634
Device Sequence Number10594969
Product Code LPH
UDI-Device Identifier00888912078665
UDI-Public00888912078665
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
K130099
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 01/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number425-97-008
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/09/2025
Initial Report FDA Received Date01/09/2026
Date Device Manufactured08/21/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
400-03-362 LOT: 866B2263
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient SexFemale
-
-