| Model Number |
V+C, ENGLISH |
| Medical Device Problem Code |
Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/10/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H6: the device was evaluated by ventec where the reported issue of it rebooting by itself was confirmed.Ventec reseated the internal flow transducer cable (ift) to resolve the reported issue.Proper device operation was then confirmed through functional and performance testing.The investigation determined that the root cause of the reported issue was that the ift cable was not fully seated.
|
| |
|
Event or Problem Description
|
|
It was reported that the device needed service.Upon evaluation of the device, ventec observed that it was rebooting by itself.There were no reports of patient use associated with the reported issue.
|
| |
|