• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION CORPORATION LINEAR? 3-6; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION CORPORATION LINEAR? 3-6; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF Back to Search Results
Model Number SC-2366-70
Medical Device Problem Codes High impedance (1291); Unexpected Therapeutic Results (1631); Migration (4003)
Health Effect - Clinical Codes Pain (1994); Discomfort (2330)
Date of Event 11/25/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional suspect medical device components involved in the event: product family: scs-linear leads.Upn: m365sc2366700.Model: sc-2366-70.Serial: (b)(6).Batch: 20088707.Udi: (b)(4).Product family: scs-linear leads.Upn: m365sc2366700.Model: sc-2366-70.Serial: (b)(6).Batch: 20019698.Udi: (b)(4).Product family: scs-linear leads.Upn: m365sc2366700.Model: sc-2366-70.Serial: (b)(6).Batch: 20019698.Udi: (b)(4).
 
Event or Problem Description
It was reported that the patient was experiencing inadequate pain relief and discomfort.Imaging confirmed that the leads migrated, high impedances were noted.The patient underwent a revision procedure wherein the leads were repositioned.The patient was doing well postoperatively.Noting was removed or added.
 
Additional Manufacturer Narrative
Investigation results: the ipg and leads were not returned for analysis; therefore, a technical analysis could not be performed.The patient underwent a revision procedure wherein the leads were repositioned.The patient was doing well postoperatively, and the leads remain implanted and in use.Device history record: it was confirmed this device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Labeling review: review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.Investigation conclusion: based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of known inherent risk of device.
 
Event or Problem Description
It was reported that the patient was experiencing inadequate pain relief and discomfort.Imaging confirmed that the leads migrated, high impedances were noted.The patient underwent a revision procedure wherein the leads were repositioned.The patient was doing well postoperatively.Nothing was removed or added.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINEAR? 3-6
Common Device Name
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION CORPORATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
25155 rye canyon loop
valencia CA 91355
Manufacturer Contact
erik sherburne
25155 rye canyon loop
valencia, CA 91355
7632920920
MDR Report Key24025577
Report Number3006630150-2026-00229
Device Sequence Number8939727
Product Code LGW
UDI-Device Identifier08714729789567
UDI-Public(01)08714729789567(17)181130(10)20019698
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P030017
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/30/2018
Device Model NumberSC-2366-70
Device Catalogue NumberSC-2366-70
Device Lot Number20019698
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/19/2025
Supplement Date Received by Manufacturer04/11/2026
Initial Report FDA Received Date01/09/2026
Supplement Report FDA Received Date05/06/2026
Date Device Manufactured11/30/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age54 YR
Patient SexFemale
-
-