| Model Number |
3789 |
| Medical Device Problem Code |
Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Code |
Inadequate Pain Relief (2388)
|
| Date of Event |
12/16/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event is estimated.The results/method and conclusion codes along with investigation results will be provided in the final report.Further information was requested but not yet received.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient did not charge their ipg as recommended and the ipg became unable to communicate with external devices.As a result, surgical intervention was undertaken wherein the ipg was explanted and replaced on (b)(6) 2025 to address the issue.During ipg replacement the patients lead had a high impedance.As a result, the lead was also replaced on (b)(6) 2025 to address the issue.
|
| |
|
Additional Manufacturer Narrative
|
|
The directions for use states that the implant must be charged every 30 days to avoid battery depletion.Based on the information received, the cause of the reported incident is related to user error.
|
| |
|