• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® THORACOABDOMINAL BRANCH ENDOPROSTHESIS; ENDOVASCULAR SYSTEM FOR TREATMENT OF THORACOABDOMINAL AND PARARENAL AORTIC LESIO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® THORACOABDOMINAL BRANCH ENDOPROSTHESIS; ENDOVASCULAR SYSTEM FOR TREATMENT OF THORACOABDOMINAL AND PARARENAL AORTIC LESIO Back to Search Results
Catalog Number ATAA43720160A
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Ischemia (1942)
Date of Event 12/12/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Updated section g3/g4, h2, and h7.Further review of this event has determined there is no allegation against the gore® excluder® thoracoabdominal branch endoprosthesis (tambe).The dissection that occurred in superior mesenteric artery (sma) was a result of a non-gore device-guide wire.Patient expired due to the complications from dissection which likely led to mesenteric ischemia.The physician stated he attributes the patient's death to complications from the dissection and mesenteric ischemia (bowel ischemia), and not to any implanted gore devices.This report shall be retracted as this is not a reportable event, and there is no allegation against the gore tambe device.
 
Event or Problem Description
The following was reported to gore: the patient had a history of undergoing a procedure in which a physician modified graft (pmeg) was implanted.It was reported a non-gore stent was also implanted the superior mesenteric artery (sma) during this procedure.It was reported the pmeg was leaking into the aneurysm.To address the aneurysm growth, plans were made to use a gore® excluder® thoracoabdominal branch endoprosthesis (tambe) to reline the existing modified graft.On (b)(6) 2025, the patient underwent the tambe procedure.A gore® excluder® thoracoabdominal branch endoprosthesis (tambe) and gore® viabahn® vbx balloon expandable endoprosthesis (vbx device) were implanted.It was reported after the tambe aortic component device was implanted, during cannulation of the sma, the physician inadvertently caused a dissection while advancing the guidewire into the existing non-gore stent.As reported, the dissection was covered with deployment of the vbx device into portal of the tambe device.The vbx device was deployed with no issues.The procedure was completed with no further issues, and the patient tolerated the procedure it was reported that on (b)(6) 2025, the patient died due to mesenteric ischemia.It was reported that at the end of the procedure, final imaging showed good results and no dissection was evident.The physician stated he attributes the patient's death to complications from the dissection and mesenteric ischemia, and not to any implanted gore devices.
 
Additional Manufacturer Narrative
H6: code c21 - results pending completion of product history review.H6: code c20 - the device remains implanted and, therefore, was not available for direct analysis by gore.H6: code a26-used to capture ischemic event with unknown cause.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® THORACOABDOMINAL BRANCH ENDOPROSTHESIS
Common Device Name
ENDOVASCULAR SYSTEM FOR TREATMENT OF THORACOABDOMINAL AND PARARENAL AORTIC LESIO
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL WOODY SPRINGS B/P
3450 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
lillian espinoza
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key24031069
Report Number2017233-2026-07035
Device Sequence Number21171536
Product Code QZK
UDI-Device Identifier00733132659685
UDI-Public(01)00733132659685(17)280614
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P230023
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 01/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberATAA43720160A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/12/2025
Supplement Date Received by Manufacturer01/12/2026
Initial Report FDA Received Date01/09/2026
Supplement Report FDA Received Date01/13/2026
Date Device Manufactured06/16/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-