• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION FINELINE II STEROX; IMPLANTABLE LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION FINELINE II STEROX; IMPLANTABLE LEAD Back to Search Results
Model Number 4457
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Erythema (1840); Unspecified Infection (1930)
Date of Event 07/11/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this implantable lead was part of the system revision due to infection and pocket redness.No adverse patient effects were reported.The implantable lead remains implanted.
 
Additional Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, additional information from product investigation indicates that the device was not returned for analysis; therefore, a technical analysis could not be performed.Good faith effort attempts were made to try and retrieve the device; however, response was not received.It was confirmed that a review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There was no indication the device manufacturing process contributed to the reported complaint.A risk review was completed and confirmed that the event of infection was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Review of labeling determined that the complaint situation was listed in the manual.There was no indication in the complaint that the product was not used in accordance with labeling.The manual was unlikely to be the cause of the reported complaint; translation, wording, or graphics does not require further review.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of known inherent risk.This supplemental report is being filed to correct the d9: device avail for eval; d9: device return date; g2: report source; h2: follow-up type; h3: device eval by manufacturer; and h11: additional mfr narrative.
 
Event or Problem Description
It was reported that this implantable lead was part of the system revision due to infection and pocket redness.No adverse patient effects were reported.The implantable lead remains implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FINELINE II STEROX
Common Device Name
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key24031939
Report Number2124215-2026-01739
Device Sequence Number10276994
Product Code DTB
UDI-Device Identifier00802526060915
UDI-Public(01)00802526060915(17)160811
Combination Product (Y/N)Y
Initial Reporter CountryUS
PMA/510(K) Number
P960004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date08/11/2016
Device Model Number4457
Device Catalogue Number4457
Device Lot Number783929
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/15/2025
Supplement Date Received by Manufacturer03/04/2026
Initial Report FDA Received Date01/12/2026
Supplement Report FDA Received Date03/24/2026
Date Device Manufactured08/11/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age70 YR
Patient SexMale
-
-