| Model Number |
NM-F4208 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
11/24/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The dhr was reviewed for lot #9001217 and there was no evidence discovered to indicate that a product defect or non-conformity contributed to the issue.The investigation for this event has not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Our manufacturing system assures, that production and process controls are in place to ensure that batches conform to the applicable specifications before they are distributed.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
|
| |
|
Event or Problem Description
|
|
The clinician reports the implant was inserted (b)(6) 2025.On (b)(6) 2025 , loosening of implant not related to bone ingrowth was verified.According to the information, the patient is healthy.The doctor said: implant did not integrate.A secondary implant was placed after removal.The device was forwarded to the manufacturer.No patient operative or post-operative complications reported.
|
| |
|
Search Alerts/Recalls
|