MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON CLAREON ASPHERIC HYDROPHOBIC ACRYLIC IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON CLAREON ASPHERIC HYDROPHOBIC ACRYLIC IOL; INTRAOCULAR LENS Back to Search Results
Model Number SY60WF
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/22/2025
Type of Reportable Event Malfunction
Event or Problem Description
A nurse reported that during an intraocular lens (iol) implant procedure, the lens was scratched.The surgery was completed on the same day with unspecified lens.Additional information has been requested.
 
Additional Manufacturer Narrative
A product was not returned for analysis.Complaint history and product history records were reviewed, and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
Additional information has been received and stated that a scratched lens was noticed after lens was implanted.The lens was removed during initial procedure and replaced with company lens.There was no patient harm.
 
Additional Manufacturer Narrative
The lens was returned positioned incorrectly in the lens case inside the lens carton.Solution was dried on the lens.One haptic/optic junction edge was broken.The missing lens material from the broken area was not returned.This haptic was also broken in the distal area (not returned).The optic was cut into two pieces, typical of an insertion and removal.The lens was cleaned with klrp for further evaluation.One optic portion had a linear scrape mark on the posterior surface near the optic center.The optic edge was torn.The anterior optic surface had a small scratch near the opposite edge.The second optic portion had a linear scrape mark on the posterior surface near the optic center which aligned with the damage of the first optic portion.The posterior surface had several small cracked damaged areas.The opposite side of this lens portion had cracked damage in the shape of an instrument used to grasp the lens, seen on the anterior and directly opposite on the posterior surface.A device history record review and a non-conformance review of the reported lot number was conducted.The deviation review did not reveal any potential contributing factors to the reported complaint and all corresponding production release specifications defined in the device master record were met.Qualified associated products were used.The reported scratch damage was observed.All lenses are 100% inspected for cosmetic attributes and the damage exhibited by the returned complaint sample would not have met company release criteria.Based on our observation, and the review of manufacturing records, it can be reasonably concluded that the damage is not manufacturing related.Due to the presence of surgical solution and the condition of the returned sample, the damage is most likely related to customer handling.The scratch was noted to be observed after the lens was implanted.The posterior of the optic has a linear scrape mark in the optic center, consistent with damage caused from a plunger underride.The returned lens was cut into two portions, typical in appearance to an insertion and removal.Additional lens damage was also observed.The ifu instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd-filled cartridge.The lens should be inserted until the optic is a little more than half-way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and/or damage.The ifu instructs to completely fill the cartridge with ovd immediately prior to loading and delivery of the lens.Do not attempt to load the lens without adequate ovd in the device.Not adequately filling the device with viscoelastic will result in inadequate coverage of lens and the lens fold path with ovd, which may result in damage.Important: the plunger should make initial contact with the cartridge at the ramp.In the event the plunger does not contact the cartridge at the ramp, do not use the handpiece and contact company.The handpiece ifu instructs: verify that the plunger tip is properly engaging the lens optic and that the lens moves forward at the same rate as the plunger while slowly advancing the plunger forward to avoid damaging the lens.When the threads on the knob make contact with the barrel, turn the knob clockwise approximately ½ turn to engage the threads and then stop.The iol will now be in the dwell position.Inspect to ensure the plunger is behind the optic.The response indicated the lens was in the cartridge for less than 3 minutes.This is within the recommendations of the ifu.The ifu instructs: follow the section regarding directions for use for information on the maximum allowed time for the iol to stay in the folded condition.Failure to adhere to manufacturer¿s recommendations may result in iol damage.Ifu note: during lens loading and insertion, do not allow the clareon iol to remain in a folded condition within the selected iol delivery system for more than 3 minutes prior to completing insertion into the capsular bag.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON ASPHERIC HYDROPHOBIC ACRYLIC IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key24035275
Report Number1119421-2026-00093
Device Sequence Number18394960
Product Code HQL
UDI-Device Identifier00380652251372
UDI-Public(01)00380652251372(17)280825(11)240827(10)15913172
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 02/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSY60WF
Device Catalogue NumberSY60WF.145
Device Lot Number15913172
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/23/2025
Supplement Date Received by Manufacturer01/15/2026
Initial Report FDA Received Date01/12/2026
Supplement Report FDA Received Date02/09/2026
Date Device Manufactured08/27/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DUOVISC OVD; MONARCH III IOL, CARTRIDGE D; MONARCH III IOL, INJECTOR
Patient Age59 YR
Patient SexFemale
-
-