• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number SRM-360
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 06/12/2025
Type of Reportable Event Serious Injury
Event or Problem Description
Healthcare professional reported "exchange with no complaint against the device".Then healthcare professional reported "rupture".This record is for the left side.Device was explanted and replaced.Device will be returned.
 
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
 
Additional Manufacturer Narrative
Additional, changed, and/or corrected data: b6, d.9, h.3, h.6.Device evaluation: per the investigation procedure, the device is analyzed through visual inspection microscopic inspection if openings are observed and a weight verification.Per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required: ¿ rupture: observed broken device through microscopic inspection assessed as fold flaw opening.No further actions are required as it is not related to the manufacturing process.None of the other observations performed during the device analysis non penetrating nicks, wear abrasion and creases are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
 
Event or Problem Description
Healthcare professional reported "exchange with no complaint against the device".Then healthcare professional reported "rupture".This record is for the left side.Device was explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS 
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS  
Manufacturer Contact
terry ingram
2525 dupont drive
irvine, CA 92612
8479366324
MDR Report Key24036007
Report Number9617229-2026-00761
Device Sequence Number21453185
Product Code FTR
UDI-Device Identifier10888628007116
UDI-Public(01)10888628007116(10)2864816(11)160201(17)210101
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2025
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/01/2021
Device Catalogue NumberSRM-360
Device Lot Number2864816
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/17/2025
Supplement Date Received by Manufacturer02/24/2026
Initial Report FDA Received Date01/12/2026
Supplement Report FDA Received Date03/11/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/01/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age52 YR
Patient SexFemale
Patient Weight69 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-