| Catalog Number |
45-700 |
| Medical Device Problem Codes |
Device Appears to Trigger Rejection (1524); Material Rupture (1546)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
11/19/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of "capsular contracture" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture and capsular contracture, baker grade iii.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "rupture" and "capsular contracture baker grade iii".This record is for left side.The device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: per the investigation procedure, the device is analyzed through visual inspection microscopic inspection if openings are observed and a weight verification.Per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required: ¿ rupture: observed an opening through microscopic inspection assessed as fold flaw opening.No further actions are required as it is not related to the manufacturing process.¿ capsular contracture: unable to observe through visual inspection as it is a physiological phenomenon.None of the other observations performed during the device analysis wear abrasion, creases and non-penetrating nicks are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "rupture" and "capsular contracture baker grade iii".This record is for left side.The device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: a2.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "rupture" and "capsular contracture baker grade iii".This record is for left side.Device was explanted.
|
| |
|