| Catalog Number |
383596 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/19/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that bd nexiva diffusics w/maxzero 22g x 1.00 in leaked.It was reported by customer that the leaking observed at connection of maxzero and diffusics.Leaking observed at connection of maxzero and diffusics.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results: our quality engineer inspected the 2 photos and 1 used physical sample submitted for evaluation.The reported issue of leakage was not confirmed upon inspection of the samples.The maxzero valve component was not provided, the luer connection was evaluated.Analysis of the sample showed that visual inspection discovered no damage was observed.A leak test was also performed and the samples showed no leakage.Bd could not determine a manufacturing related root cause since the reported defect was not confirmed during the evaluation of the sample.A review of our risk management documentation was performed and indicates that the potential risk of the reported event was assessed appropriately.Production records were reviewed, and this batch meets our manufacturing specification requirements.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
No additional information.
|
| |
|