ALCON LABORATORIES IRELAND LTD. CLAREON IOL WITH THE AUTONOME DELIVERY SYSTEM; INTRAOCULAR LENS Back to Search Results
Model Number CNA0T0
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/16/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A non-healthcare professional reported during intraocular lens (iol) implant procedure, the lens would not discharge, the plunger slide under lens with patient contact and no patient harm.The lens remained in the delivery system.The procedure was completed on same day.Additional information has been requested.
 
Additional Manufacturer Narrative
Additional information was provided in d.9., h.3., h.6.And h.11.The product was returned for analysis and the reported complaint was observed.The device was returned loose in a plastic bag.The lock-out assembly has been removed.Viscoelastic is observed in the device.The plunger has been fully advanced outside the nozzle tip and is advanced under the intraocular lens (iol).The iol is advanced into the nozzle entry and is damaged.The plunger tip is bent/damaged, this could possibly be due to the device being returned loose in a plastic bag.We are unable to determine the root cause for the reported complaint "would not discharge, plunger slide under lens, patient contact with no harm".Plunger underride was observed.Plunger underride may occur: if the lens has become misaligned in the lens bay during the manufacturing process.Due to the use of a non-qualified viscoelastic.Material properties of non-qualified ophthalmic viscosurgical device (ovd) may contribute to underfill, overfill, misfolding of the haptics, or other inconsistent folding outcomes.If inadequate viscoelastic is placed in the device this will cause inadequate coverage of the lens fold path which may cause the lens to advance incorrectly or become "stuck" in the device allowing the plunger to underride the lens.If the operating room temperature is too high (greater than 23 degrees centigrade / 73 degrees fahrenheit) lens folding consistency is negatively affected, the lens is more adherent and this may inhibit lens advancement.Any of the above listed causes alone, or in combination, may create the reported event.Based on the results from the product history record, the products met release criteria.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON IOL WITH THE AUTONOME DELIVERY SYSTEM
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key24049988
Report Number9612169-2026-00116
Device Sequence Number18578275
Product Code HQL
UDI-Device Identifier00380652393676
UDI-Public(01)00380652393676(17)280427(11)250429(10)26020593
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCNA0T0
Device Lot Number26020593
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/19/2025
Supplement Date Received by Manufacturer03/18/2026
Initial Report FDA Received Date01/12/2026
Supplement Report FDA Received Date04/10/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/29/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DUOVISC VISCOELASTIC SYSTEM.
Patient Age75 YR
Patient SexFemale
-
-