MAUDE Adverse Event Report: EDWARDS LIFESCIENCES EDWARDS INSPIRIS RESILIA AORTIC VALVE; HEART-VALVE, NON-ALLOGRAFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES EDWARDS INSPIRIS RESILIA AORTIC VALVE; HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Model Number 11500A
Medical Device Problem Codes Degraded (1153); Device Stenosis (4066)
Health Effect - Clinical Codes Heart Failure/Congestive Heart Failure (4446); Insufficient Information (4580)
Date of Event 12/17/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient with a 21mm 11500a aortic valve underwent a valve-in-valve procedure after an implant duration of 7 years, 5 months due to moderate stenosis.The procedure was performed with a 23mm 9755rsl transcatheter valve.The procedure was completed successfully.
 
Additional Manufacturer Narrative
The subject device is not available for evaluation as it remains implanted in the patient.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Additional Manufacturer Narrative
Updated sections b4, b5, b7.Updated sections g3, g6.Updated sections h2, h6 - investigation findings; investigation conclusions.H11: additional manufacturer narrative: svd is a logical and expected consequence of the chemical, mechanical, and immunological processes associated with bhv implantation.The common endpoint of all these factors is calcification and degradation.Svd is an acquired condition, intrinsic to bioprosthetic valves, characterized by deterioration of the leaflets or supporting structures, leading to thickening, calcification, tearing, or disruption of the prosthetic valve materials.These changes result in valvular dysfunction manifesting as stenosis and/or regurgitation with a consequent drop in hemodynamic efficacy of the valve.Despite the advancements in the heart valve industry, svd commonly begins 7- 8 years after implantation, with a considerably increased rate in patients with pertinent comorbidities and/or an implant duration of greater than 10 years.Prior investigations and extensive literature review have shown that svd is predominantly related to patient factors and implant duration rather than to manufacturing issues.Events related to device design, manufacturing, or use error tend to manifest at an earlier implant duration.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.Based on the information available, the most likely root cause is patient factors, including an implant duration of 7 or more years, hyperlipidemia (hld), and coronary artery disease (cad).The subject device is not available for evaluation as it remains implanted in the patient.The instructions for use (ifu) have been reviewed, and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.The reported type of event is included in the ifu.The device history record (dhr) review was completed, and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Event or Problem Description
It was reported that a patient with a 21mm 11500a aortic valve underwent a valve-in-valve procedure after an implant duration of 7 years, 5 months due to degeneration and moderate stenosis.The patient presented with heart failure.The tavr procedure was successfully performed with a 23mm 9755rsl valve.The patient was stable post-procedure and was discharged on pod #1.
 
Event or Problem Description
It was reported that a patient with a 21mm 11500a aortic valve underwent a valve-in-valve procedure after an implant duration of 7 years, 5 months due to degeneration and moderate stenosis.The patient presented with symptoms of heart failure.The procedure was performed with a 23mm 9755rsl transcatheter valve.The procedure was completed successfully.Patient was stable post-procedure and was discharged on pod# 1.
 
Additional Manufacturer Narrative
H11: additional manufacturer narrative updated sections b4, b5, b7.Updated sections g3, g6.Updated sections h2, h6 - health effect - clinical code.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.H11: corrected data: corrected section h6 - device code(s).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS INSPIRIS RESILIA AORTIC VALVE
Common Device Name
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer Contact
jason reves
1 edwards way
mle fl2
irvine, CA 92614
9492503859
MDR Report Key24050384
Report Number2015691-2026-10374
Device Sequence Number9855807
Product Code LWR
UDI-Device Identifier00690103194975
UDI-Public(01)00690103194975(17)190915
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
P150048
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/04/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/15/2019
Device Model Number11500A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/17/2025
Supplement Date Received by Manufacturer01/13/2026
02/27/2026
Initial Report FDA Received Date01/13/2026
Supplement Report FDA Received Date02/10/2026
03/04/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/15/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Life Threatening; Required Intervention;
Patient Age75 YR
Patient SexMale
-
-