• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CRITICAL CARE AB BASE UNIT SERVO-N; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CRITICAL CARE AB BASE UNIT SERVO-N; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 6688600
Medical Device Problem Code Failure of Device to Self-Test (2937)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/02/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the ventilator failed pressure transducer test during pre-use check.There was no patient involvement.Manufacturer's ref.#: (b)(4).
 
Additional Manufacturer Narrative
On-site investigation was performed by our field service engineer.The claimed issue was reproduced during troubleshooting.According to received information in service report, the replacement of the pressure transducer printed circuit board (pcb) solved the issue.The ventilator passed all functional and safety tests and was cleared for clinical use.Pressure transducer pcb measures the pressure, conveyed via the pressure tube connected to this block by its differential pressure transducer.With differential reference to the ambient pressure, the output signal is proportional to the measured pressure thus giving a linear measurement in the range -40 cmh2o to +160 cmh2o.The device's logs were not provided for further analysis.The cause of the claimed issue in this customer product complaint has been determined.The issue was related to pressure transducer pcb.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BASE UNIT SERVO-N
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer (Section G)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer Contact
caroline kabbabe
roentgenvagen 2
solna 
MDR Report Key24050657
Report Number8010042-2026-0000024
Device Sequence Number13225524
Product Code CBK
UDI-Device Identifier07325710001042
UDI-Public(01)07325710001042(11)250318
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K201874
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number6688600
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/30/2025
Initial Report FDA Received Date01/13/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/18/2025
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-