| Model Number |
453956 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Combination product: yes.
|
| |
|
Event or Problem Description
|
|
The orsiro mission drug-eluting stent system was selected for treatment of a severely calcified lesion (90% stenosis degree) in a moderately tortuous segment of the proximal rca.After pre-dilatation and using a guideliner, the orsiro mission could not cross the lesion.The stent frayed upon removal.
|
| |
|
Additional Manufacturer Narrative
|
|
Combination product: yes.Neither the affected device nor the angiographic material was returned.Therefore, no technical investigation on the subject could be performed.The production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the production documentation confirmed that the device was manufactured according to specifications and passed all in-process and final inspections.During final inspection, every stent system undergoes visual inspection to ensure correct stent embedding and homogeneous crimping of the stent.Based on the conducted investigations, no material or manufacturing related root cause could be determined.Considering the event description provided, the most probable root cause for the reported event is related to the patient's anatomy (i.E.Severe calcification).
|
| |
|
Event or Problem Description
|
|
The orsiro mission drug-eluting stent system was selected for treatment of a severely calcified lesion (90% stenosis degree) in a moderately tortuous segment of the proximal rca.After pre-dilatation and using a guideliner, the orsiro mission could not cross the lesion.The stent frayed upon removal.
|
| |
|