| Model Number |
MI1250 SYNCHRONY 2 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
The device was explanted due to electrode migration from the cochlea.Ct scan was ordered and confirmed the migration.The user was re-implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: device investigation did not reveal any device defect or damage, which has been present whilst implanted.Mechanical damages found are attributable to the removal surgery.According to the information received from the field the device was explanted due to a migration of the active electrode out of cochlea.Further one channel was already left extra-cochlear at implantation surgery.This is a final report.
|
| |
|
Event or Problem Description
|
|
The device was explanted due to electrode migration from the cochlea.Ct scan was ordered and confirmed the migration.The user was re-implanted.
|
| |
|