• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. PERSONA FEM CR POR CCR NRW SZ 8 L; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. PERSONA FEM CR POR CCR NRW SZ 8 L; PROSTHESIS, KNEE Back to Search Results
Catalog Number 42502206401
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pain (1994)
Date of Event 12/22/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D10: 42540200032 - all poly pat ve 32 mm dia - 65061520.42532007101 - psn tib stm 5 deg sz e l - 65361036.42512100810 - psn mc ve asf l 10mm 8-11/ef - 65362700.66022663 - palacos r + g (1x40) - 60771089.G2: australia.H6: mechanical (g04) - femur.Customer has indicated that the product will not be returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the patient underwent an initial left total knee arthroplasty.Approximately 3.5 years post-op, the patient was experiencing pain without relief from physiotherapy, a pain management specialist, and rehabilitation.Imaging revealed increased uptake at the patella and medial tibial area.Subsequently, the patient was revised.All components were revised.There was no loosening seen, however, the surgeon noted areas of softer bone.Attempts have been made and all available information has been provided.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; d2; g1; g3; g6; h1; h2; h3; h4; h6; h10.The event cannot be confirmed.Visual examination of the provided pictures showed the explanted implants; visual and dimensional evaluations could not be performed.The device history record was reviewed and no discrepancies relevant to the reported event were found.Review of complaint history found no additional related issues for this item and the reported part and lot combination.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: radiolucency underlying the medial tibial plate is suspicious for osteolysis and a low lying patella.These findings may be associated with pain and increased activity on bone scan.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusion or information, a supplemental report will be filed accordingly, zimmer biomer will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSONA FEM CR POR CCR NRW SZ 8 L
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key24052775
Report Number0001822565-2026-00151
Device Sequence Number15916654
Product Code MBH
UDI-Device Identifier00889024470095
UDI-Public(01)00889024470095(17)320417(10)65313223
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K172524
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue Number42502206401
Device Lot Number65313223
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/22/2025
Supplement Date Received by Manufacturer01/23/2026
Initial Report FDA Received Date01/13/2026
Supplement Report FDA Received Date01/28/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/20/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-