Model Number DISSECTOR, CURVED, HL 4.2MM X 19CM
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/19/2025
Type of Reportable Event Malfunction
Event or Problem Description
On 19-dec-2025, it was reported by a sales representative via (b)(4) that an ar-6420cds curved dissector left metal shavings behind.This was discovered during the case with no patient harm.Additional information was provided by the rep on 23-dec-25 that this was discovered during a hip scope.The shavings were discovered during the debridement of labrum.All fragments were retrieved.Both suction and irrigation were used, and no alarms were present.Another device was used to complete the case.
 
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.-complaint allegation is not confirmed.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause is attributed to misuse of the device, including exposure to excessive mechanical forces, misaligned insertion, and/or side loading during use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISSECTOR, CURVED, HL 4.2MM X 19CM
Common Device Name
SAW BLADE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key24053050
Report Number1220246-2026-00186
Device Sequence Number9883594
Product Code GFA
UDI-Device Identifier00888867039193
UDI-Public(01)00888867039193(11)250828(17)300731(10)15489812
Combination Product (Y/N)N
Initial Reporter StateSD
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 04/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDISSECTOR, CURVED, HL 4.2MM X 19CM
Device Catalogue NumberAR-6420CDS
Device Lot Number15489812
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/19/2025
Supplement Date Received by Manufacturer04/20/2026
Initial Report FDA Received Date01/13/2026
Supplement Report FDA Received Date04/22/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/28/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-