| Model Number |
DISSECTOR, CURVED, HL 4.2MM X 19CM |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/19/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On 19-dec-2025, it was reported by a sales representative via (b)(4) that an ar-6420cds curved dissector left metal shavings behind.This was discovered during the case with no patient harm.Additional information was provided by the rep on 23-dec-25 that this was discovered during a hip scope.The shavings were discovered during the debridement of labrum.All fragments were retrieved.Both suction and irrigation were used, and no alarms were present.Another device was used to complete the case.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.-complaint allegation is not confirmed.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause is attributed to misuse of the device, including exposure to excessive mechanical forces, misaligned insertion, and/or side loading during use.
|
| |
|