| Model Number |
SIGSDS30CTV |
| Medical Device Problem Code |
Misfire (2532)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/23/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter, intra-operatively of an open liver resection, during vascular treatment, the stapler was stuck after a single firing grip.The firing was stopped, and the device was withdrawn using the black return knob.It was reported that the firing had been partially completed up to the point where the handle was gripped.Afterwards, a different cartridge was attached, the firing was performed, and the procedure was completed without further incident.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
D10 concomitant product: egiaustnd, egiaustnd endogia ultra univ std stap (lot#p5e0904); unk - sinh, unknown surgical innovations nh product (lot#unknown); sigpshell, sig power sigpshell control shell (lot#unknown).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the reload was partially fired with the interlock engaged.The jaws were open.Staple pushers were visible at the cut line.Functional testing noted that the reload was loaded into a representative instrument.The interlock was overridden and the reload was applied to test media.All remaining staples were placed, and test media was cleanly transected.The reload interlock was tested and found to function properly.It was reported that the device experienced partial firing.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur if the firing handle is not fully cycled.If the instrument return knobs are retracted at any point once the firing cycle has begun, the reload will engage into safety interlock and prevent further attempts to fire by ceasing the placement of staples and tissue transection and prevent patient harm.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: failure to completely fire the reload will result in an incomplete cut or incomplete staple formation, which may result in poor hemostasis or leakage.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|