| Model Number |
DDU-100B |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Although requested, the log files from the device have not been provided and the cause of the complaint is not known.
|
| |
|
Event or Problem Description
|
|
A customer reported their aed is flashing red and prompting a "replace battery pack now" warning.They reported this did not occur during patient use.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis of the log files from the aed identified that the customer was performing excessive self-testing of the aed which reduced the battery life expectancy.As a follow-up with the customer, the proper maintenance of this device was reviewed.
|
| |
|
Additional Manufacturer Narrative
|
|
Troubleshooting of the aed with the customer identified that once the new battery was installed and a manual self test run the aed was functioning as designed and could stay in service.Troubleshooting of the aed with the customer did not identify a cause for the depleted battery pack.As a follow up with the customer, the proper maintenance of this device was reviewed.
|
| |
|