| Catalog Number |
81324 |
| Medical Device Problem Code |
Restricted Flow rate (1248)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
While using a cavitron g124 generator they allege that they have low water flow and the handpiece is getting hot, no injury resulted.
|
| |
|
Additional Manufacturer Narrative
|
|
There has been a previous report received where lack of water flow has caused an overheating insert.Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
| |
|
Additional Manufacturer Narrative
|
|
01/21/26 evaluation tech: (b)(4).Emv hp; cable, conn/gun cable open.Blockage in the handpiece cable causing restricted water flow.Debris buildup in the water filter.Cabinet is discolored.Shorted g123 harness wire.Pump not holding air pressure.Will replace damaged/worn components and recalibrate unit to factory specs upon estimate approval.No water hose, handpiece, power cord, power supply received for evaluation.Hpc-04150, st-n/a.Dhr review is not required because the product was returned for evaluation, and the described failure mode is a known hazard.
|
| |
|