• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH 2.8MM INTERLOCKING BOLT FOR PERCUTANEOUS INSERTION HANDLE; GUIDE, SURGICAL, INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH 2.8MM INTERLOCKING BOLT FOR PERCUTANEOUS INSERTION HANDLE; GUIDE, SURGICAL, INSTRUMENT Back to Search Results
Catalog Number 03.124.006
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/31/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4), this report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: h3, h6: product investigation summary: the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the threaded tip of the 2.8 interlckng bolt for perc insrt h&le was worn/stripped.The observed condition was consistent as an end-of-life indicator for the device.The observed condition was identified as an end-of-life indicator, damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.A dimensional inspection was not performed due to post manufacturing damage.An interaction test was performed with the mating plate, and a difficulty was detected in correctly assembling the threaded tip into the holes of the plate.Damage to the bolt threads prevents proper connection with the plate in some of the locking holes.The overall complaint was confirmed as the observed condition of 2.8 interlckng bolt for perc insrt h&le would have contributed to the complained issue.It was determined that the reusable instrument was worn from repeated use and servicing.Based on the investigation findings, the potential cause is traced to end of life, and it has been determined that no corrective and/or preventative action is required.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.H3, h4, h6: device history record (dhr) review conducted: part# 03.124.006 lot # 7907487 manufacturing site: werk hagendorf manufacturing date: 21.June.2012 a manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that during a procedure on (b)(6) 2025, the tibial plateau plate hole and interlocking nut stripped.Another set was opened to complete the procedure.There was no generation of fragments.There was no surgical delay.There was no patient consequence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.8MM INTERLOCKING BOLT FOR PERCUTANEOUS INSERTION HANDLE
Common Device Name
GUIDE, SURGICAL, INSTRUMENT
Manufacturer (Section D)
SYNTHES GMBH
luzernstrasse 19-21
zuchwil 4528
SZ  4528
Manufacturer (Section G)
SYNTHES HAEGENDORF GMBH
im bifang 6
haegendorf 4614
SZ   4614
Manufacturer Contact
elba bello
1302 wrights lane east
west chester, PA 19380
9083863534
MDR Report Key24056978
Report Number8030965-2026-00446
Device Sequence Number21303130
Product Code FZX
UDI-Device Identifier10886982075000
UDI-Public(01)10886982075000(10)7907487
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.124.006
Device Lot Number7907487
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/09/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/29/2025
Initial Report FDA Received Date01/13/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/21/2012
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-