| Model Number |
TA9048S |
| Medical Device Problem Code |
Failure to Fire (2610)
|
| Health Effect - Clinical Code |
Failure to Anastomose (1028)
|
| Date of Event |
12/15/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D10 concomitant product: ta9048s, ta9048s ta 90-4.8 reloadable stapler (lot # p4j0779) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that intraoperatively, a small bowel fistula due to an incisional hernia plaque via laparotomy.To perform the anastomosis, the surgeon first created a lateral anastomosis of the small bowel using a gia stapler, then a lateral anastomosis using a ta stapler.The initial ta stapler did not fire, no staples were delivered.A second ta stapler was retrieved.This time, the stapler was closed on the small bowel, but only a few staples were placed, and the incision was not completely closed.The anastomosis was not performed under the most aseptic conditions possible.There was injury to the edges of the anastomosis, a cut in the small bowel, and consequent shortening of the anastomosis.Two additional staplers from another company were used to complete the procedure successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the initial ta stapler did not fire.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|