BOSTON SCIENTIFIC CORPORATION AXIOS?; UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) Back to Search Results
Model Number M00553680
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/01/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
D4, h4: the complainant was unable to provide the complaint device lot number; therefore, the device manufacture date is unknown at this time.If additional details become available, a supplemental report will be submitted.H6: imdrf device code a150201 captures the reportable event of stent failure to deploy.H11: boston scientific is initiating a removal of certain sizes of the axios stent and electrocautery enhanced delivery system (6mm x 8mm, 8mm x 8mm and 20mm x 10mm) manufactured since march 1, 2025.This action is being taken due to increased reports of stent deployment and expansion issues with these configurations.These issues only occur at the time of stent delivery and are expected to be noticed by the physician.Patients who have been treated with a successfully implanted axios stent should continue to follow standard of care and are not affected by this issue.A letter was sent out to materials managers/health care professionals on 19-dec-2025 to immediately stop further distribution or use of any affected 6mm x 8mm, 8mm x 8mm and 20mm x 10mm axios stent and electrocautery enhanced delivery system.The letter indicates to remove these devices from inventory and segregate them in a secure location until they can be returned to boston scientific.Despite multiple efforts, bsc was unable to confirm the batch number for the axios stent and electrocautery enhanced delivery system used in this event.Based on the reported clinical observations associated with failure to deploy and the date bsc was informed of this event, bsc is conservatively associating this event with the recall in an abundance of caution.The referenced axios stent and electrocautery enhanced delivery system met all specifications prior to final approval for distributions/sale.
 
Event or Problem Description
It was reported to boston scientific corporation that an axios stent and electrocautery enhanced delivery system was intended to be implanted in the pancreas to facilitate a pseudocyst drainage during an endoscopic ultrasound (eus) procedure performed on an unknown date.During the procedure, the distal flange did not open, it failed to deploy.The procedure was completed by replacing and using another of the same device.There were no patient complications reported as a result of this event.It was reported that the patient fully recovered.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXIOS?
Common Device Name
UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
christina lam
300 boston scientific way
marlborough, MA 01752
9145716918
MDR Report Key24058182
Report Number3005099803-2026-00189
Device Sequence Number10023150
Product Code KNS
UDI-Device Identifier00191506008086
UDI-Public(01)00191506008086
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K233318
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 01/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00553680
Device Catalogue NumberM00553680
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/18/2025
Initial Report FDA Received Date01/13/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
3005099803-12192025-004-
Patient Sequence Number1
Patient SexUnknown
-
-