| Catalog Number |
362753 |
| Medical Device Problem Code |
Device Ingredient or Reagent Problem (2910)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/16/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported while using bd vacutainer® cpt¿ cell preparation tube with sodium heparin, an unspecified number of tubes exhibited poor barrier separation after processing.There was no health impact or consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
E1: initial reporter state: address information was not able to be obtained, therefore, (b)(6) was used as a place holder.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: bd did not receive any samples or photos for investigation.A total of 60 retained samples were visually inspected, and no issues were identified.Complaints for sample quality are under statistical control for the month of december 2025.At this time, further testing is not indicated.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has not been confirmed for the indicated failure mode: poor barrier separation.No definitive root cause could be established for the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.Bd quality will continue to monitor sample quality complaints.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd vacutainer® cpt¿ cell preparation tube with sodium heparin, an unspecified number of tubes exhibited poor barrier separation after processing.There was no health impact or consequences reported.
|
| |
|
Search Alerts/Recalls
|