| Catalog Number |
363048 |
| Medical Device Problem Code |
Overfill (2404)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/18/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported while using bd vacutainer® plus citrate tube, an unspecified number of tubes overfilled.There was no health impact or consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
B3: date of event is unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd vacutainer® plus citrate tube, an unspecified number of tubes overfilled.There was no health impact or consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: bd received 2 photos for investigation, and both images show an acceptable fill level.Sufficient volume is achieved when the blood draw falls above the minimum fill indicator and below the bottom of the shield.Functional tests were conducted on 20 retained samples and were found to be within specification.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has not been confirmed for the indicated failure mode: overfill complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.
|
| |
|