| Model Number |
OMS-T10BT |
| Medical Device Problem Codes |
Break (1069); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, intra-operatively of a laparoscopic procedure, the balloon physically ruptured.A balloon fragment came out of the port insertion site during wound closure.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6 h3.Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the balloon was cut and broken.A piece was disengaged and received in a bag.During evaluation, the balloon could not be inflated due to the damage in which it was received.It was reported that the balloon physically ruptured.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The broken balloon may occur when contact is made with a surgical instrument during clinical application.While the broken lock collar may occur when excessive force is applied during clinical application.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: care must be exercised during insertion of the balloon as well as during the course of the procedure to avoid damaging the balloon with other instruments.Additionally, the investigation detected the unreported condition of broken lock collar that has no relationship to the reported condition.It was determined that the analysis noted evidence that the device was not used as intended.Damage to the balloon by instruments used during insertion and in the course of a procedure may cause device failure.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|