MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number 3505590BC
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 12/04/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: breast ptosis and capsular contracture.D6a.Implantation date: 2019.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a female patient underwent a breast augmentation revision with two 590cc mentor memorygel breast implants and experienced bilateral breast ptosis and capsular contracture (baker grade unknown) on the left side post-operatively.As a result, the patient underwent explantation on (b)(6) 2025 for the left side and replaced with a 590cc mentor memorygel breast implant.The right side was not explanted; it was addressed by surgical intervention only.This report is for the left side device.
 
Additional Manufacturer Narrative
On march 07, 2026, a photo evaluation for the device was completed.Device photo evaluation summary: upon visual evaluation of the image provided in the complaint no apparent damage or visual anomalies were observed.As the product involved in this complaint was not received, the device could not be analyzed according to our procedures.Regarding the reported capsular contracture condition, there are not enough details to determine the factors that may have caused this condition.Capsular contracture in the patient¿s breast might be the result of the body¿s individual physiological response to the implantation of a foreign object in soft tissue.Further, capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria: since no malfunction was observed during the investigation.No corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
henry boden
3041 skyway circle north
irving, TX 75038
9497882078
MDR Report Key24059696
Report Number1645337-2026-00498
Device Sequence Number9707237
Product Code FTR
UDI-Device Identifier00081317006992
UDI-Public(01)00081317006992(17)230909(10)7629624
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/09/2023
Device Catalogue Number3505590BC
Device Lot Number7629624
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/17/2025
Supplement Date Received by Manufacturer03/07/2026
Initial Report FDA Received Date01/13/2026
Supplement Report FDA Received Date03/09/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/11/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age39 YR
Patient SexFemale
-
-