• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MEXICO S. DE R.L. DE CV MARKSMAN; CATHETER, CONTINUOUS FLUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MEXICO S. DE R.L. DE CV MARKSMAN; CATHETER, CONTINUOUS FLUSH Back to Search Results
Model Number FA-55150-1030
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Codes Intracranial Hemorrhage (1891); Ruptured Aneurysm (4436)
Date of Event 12/18/2025
Type of Reportable Event Serious Injury
Event or Problem Description
Medtronic received a report that the patient had multiple intracranial aneurysms of the right internal carotid artery, including a large 10mm aneurysm at the posterior communicating artery segment and a small 4mm aneurysm at the ophthalmic artery segment.The operator planned to implant a pipeline for treatment.During the procedure to establish access, it was found that the microcatheter could not enter the lumen of the navien due to significant resistance.After replacing with a new one, it went through successfully.The marksman microcatheter was positioned, and a ped-450-30 was delivered to the distal end of the microcatheter.However, the stent could not be came out despite being able to be retrieved.Multiple attempts were made withincreased pushing force, but the problem persisted.Subsequently, possibly due to repeated forceful manipulation, the aneurysm suddenly ruptured during the procedure.The operator stated that the stent had a quality issue and would not be charged for it.The device was stuck (locked up) in the catheter/resistance in the catheter.The catheter was flushed continuously with heparanized saline.The pipeline became stuck in the distal section of the catheter during delivery.The physician released the load (slack) in the system in an attempt to resolve the issue; however this did not resolve the issue.There was no damage to the catheter or pushwire.Patient symptoms or complications associated with this event were blood vessel rupture and bleeding.The pipeline was used for an approved (on-label) indication.The device and any accessories were prepared as indicated in the ifu.The catheter was flushed as indicated in the ifu.The patient was undergoing surgery for treatment of a saccular, unruptured aneurysms on the right internal carotid artery: one at the posterior communicating artery segment and one at the ophthalmic artery segment with a max diameter of 10 mm anda 9 mm neck diameter.The landing zone was 2.5 mm distal and 4.7 mm proximal.The access vessel was the femoral artery with a diameter of 10 mm.It was noted the patient's vessel tortuosity was severe.Dual antiplatelet therapy (dapt) was administered, platelet reactivity units (pru) level was normal.The angiographic result post procedure that due to elevated intracranial pressure, the contrast agent could not go far.Ancillary devices include a navien 6f guide catheter, marksman microcatheter.Additional information was received reporting that the actions or interventions taken to resolve the ruptured aneurysm/bleeding included promptly filled spring coils, drilled holes for drainage, and relieved pressure.The procedure was completed with surgical intervention with drilling hole drainage and decompression.The cause of the resistance and rupture/bleed was unknown.It was unknown if the rupture/bleed was caused by or related to the resistance.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MARKSMAN
Common Device Name
CATHETER, CONTINUOUS FLUSH
Manufacturer (Section D)
MEDTRONIC MEXICO S. DE R.L. DE CV
av. paseo del cucapah #10510
tijuana,bc 22210
MX  22210
Manufacturer (Section G)
MEDTRONIC MEXICO S. DE R.L. DE CV
av. paseo del cucapah #10510
tijuana,bc 22210
MX   22210
Manufacturer Contact
justin ellis
9775 toledo way
irvine, CA 92618
7635265677
MDR Report Key24060658
Report Number9617601-2026-00267
Device Sequence Number15826426
Product Code KRA
UDI-Device Identifier00847536026261
UDI-Public(01)00847536026261(17)280611(10)231428268
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K091559
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFA-55150-1030
Device Catalogue NumberFA-55150-1030
Device Lot Number231428268
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/18/2025
Initial Report FDA Received Date01/13/2026
Date Device Manufactured06/11/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age81 YR
Patient SexFemale
-
-