• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INGEVITY? MRI; IMPLANTABLE LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INGEVITY? MRI; IMPLANTABLE LEAD Back to Search Results
Model Number 7732
Medical Device Problem Codes Signal Artifact/Noise (1036); Fracture (1260); High impedance (1291)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/18/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during a routine checkup, this right ventricular (rv) lead exhibited high out of range pacing impedance measurements which triggered a lead safety switch (lss) to unipolar pacing.The patient was sent to the emergency room (er) and there was noise present on the rv lead in both unipolar and bipolar pacing.Additionally, the lead was believed to be fractured.This rv lead was surgically abandoned, and no additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
If pertinent information is provided in the future, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that during a routine checkup, this right ventricular (rv) lead exhibited high out of range pacing impedance measurements which triggered a lead safety switch (lss) to unipolar pacing.The patient was sent to the emergency room (er) and there was noise present on the rv lead in both unipolar and bipolar pacing.Additionally, the lead was believed to be fractured.This rv lead was surgically abandoned, and no additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
This device was surgically abandoned and was not returned.As such, physical analysis has not been conducted in our laboratory.Based on the available information, and in the absence of a product return, the investigation conclusion code cause not established was selected.As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.A risk review was completed and confirmed that the event of fracture was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Review of labeling determined that the complaint situation was listed in the manual.There was no indication in the complaint that the product was not used in accordance with labeling.The manual was unlikely to be the cause of the reported complaint; translation, wording, or graphics does not require further review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INGEVITY? MRI
Common Device Name
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key24063347
Report Number2124215-2026-02291
Device Sequence Number10281913
Product Code NVN
UDI-Device Identifier00802526523335
UDI-Public(01)00802526523335(17)210912
Combination Product (Y/N)Y
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P150012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/05/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/12/2021
Device Model Number7732
Device Catalogue Number7732
Device Lot Number1011326
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/18/2025
Supplement Date Received by Manufacturer05/04/2026
Initial Report FDA Received Date01/13/2026
Supplement Report FDA Received Date05/05/2026
Date Device Manufactured09/12/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age79 YR
Patient SexMale
-
-