• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LUMENIS LTD SLIMLINE SIS EZ; POWERED LASER SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LUMENIS LTD SLIMLINE SIS EZ; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number 0644-020-01
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/29/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during preparation of the procedure, the fiber burned out and damaged the debris shield.The procedure was completed using another device without patient complications.Analysis of the returned device identified a fractured connector.
 
Additional Manufacturer Narrative
Upon receipt of this fiber at our quality assurance laboratory, this device was thoroughly analyzed.Analysis of the returned device confirmed the reported clinical observation as the analysis identified a connector fracture.Visual analysis did not identify any abnormality on the fiber body.Microscopic inspection of the fiber tip indicated it was degraded.Laser fibers are consumable devices and expected to degrade with use.Analysis identified the light exiting the connector face was distorted, indicating a connector fracture.There were burn marks identified on the connector face.The functional testing was performed, and the aiming beam light was dim.The following message was displayed: treatment used: 0 treatment left: 1.The observed condition of distorted light exiting the connector can be a result of an internal connector fracture.Fracture within the connector can occur during procedural handling of the fiber during setup, use or reprocessing.This connector fracture can result in an insufficient aiming beam or low laser energy output, up to a complete inability for the fiber to fire.The interaction of the console with the connector having this fracture likely led it to overheat causing the burn marks observed.The review completed on the device history review (dhr) for this device confirmed that it met specification prior to release.The ifu states, inspect the fiber for kinks, punctures, fractures, or other damage.If the fiber appears damaged, do not use the fiber, return it to the supplier for replacement.Hold the fiber tip to a non-reflective surface and ensure that a circular red/green spot appears.If the spot is not circular, cleave the fiber.If the spot is weak or not visible, discard the fiber or return it to the supplier for replacement.Based on the information available, a conclusion code of cause traced to component failure was assigned to this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SLIMLINE SIS EZ
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
LUMENIS LTD
yokneam industrial park
hakidma 6,
yokneam 20692 04
IS  2069204
Manufacturer (Section G)
MOLEX LLC
18019 n 25th ave
phoenix AZ 85023
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key24064022
Report Number2124215-2026-02329
Device Sequence Number10266727
Product Code GEX
UDI-Device Identifier17290109145669
UDI-Public(01)17290109145669(11)240727(17)290626(10)18720724
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K170121
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Distributor
Initial Reporter Occupation 505
Type of Report Initial
Report Date (Section B) 01/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0644-020-01
Device Catalogue Number0644-020-01
Device Lot Number0018720724
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/13/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/12/2026
Initial Report FDA Received Date01/13/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/27/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age43 YR
Patient SexMale
Patient Weight52 KG
-
-