A physician reported that during vitrectomy surgery, after inserting the vit probe into the eye, intermittent incomplete cutting was observed.The vit probe was replaced, and the procedure was completed successfully.There was no impact to the patient.
|
|
Additional information provided in d.9., h.3., h.6.And h.11.No technical inquiries were received from the customer.One opened probe with tip protector in a tray was received for the report.The sample was visually inspected and found to be conforming.The sample was then functionally tested for actuation and cut.The sample was found to be conforming for both tests.The probe engine was disassembled, and the components inspected.The diaphragm, spacer, top o-ring are all intact.Bottom o-ring has outer and inner diameter damage.Top o-ring has inner and outer diameter damage.A review of the device history record traceable to the lot number obtained from the device¿s radio frequency identification tag, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria the investigation conducted a non-conformance review of the reported lot number.No deviations were identified, and all corresponding production release specifications defined in the device master record were met.Based on the evaluation of the information and materials received, the investigation concluded that the root cause of the reported event is not determined as the returned sample was conforming for all functional testing.However, a manufacturing related potential contributor factor was identified, damaged was observed on the o-ring and cannot be ruled out for the actuation problem as reported.The returned sample functionally met specifications: however, a non-conformance was completed for the damage observed on the o-ring.All probes are 100% visually inspected and tested for actuation, aspiration, and cut during manufacturing.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.Complaint data for all company products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
|