| Model Number |
SUTURE ANCH BIOCOMP CORKSCREW FT VENTED |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/16/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
It was reported that during an arthroscopic knee surgery the corkscrew anchor broke.All fragments were retrieved from the patient.Per complaint reporter there was no harm for patient, operator or third party.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.One unpackaged ar-1927bcnf-1 corkscrew® (batch: 15473077) was received for evaluation.Visual inspection identified that the anchor was fractured.Functional testing was not performed due to the condition of the returned device.The most likely cause for the reported failure can be attributed to use-related factors, such as misaligned insertion, improper bone preparation, or prying/leverage applied with excessive force.The complaint allegation is confirmed.Refer to the investigation photos.
|
| |
|