| Catalog Number |
G310SCLR |
| Medical Device Problem Code |
No Flow (2991)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
While using a cavitron 300 series g310 they allege that they have no water and the handpiece is getting warm, no injury resulted.
|
| |
|
Additional Manufacturer Narrative
|
|
There has been a previous report received where lack of water flow has caused an overheating insert.Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
| |
|
Additional Manufacturer Narrative
|
|
01/23/26 evaluation tech: (b)(4).W4d water sol, iso valve build up/debris.Debris buildup in the water solenoid, debris buildup in the water supply hose, debris buildup in the water filter.Debris buildup in the handpiece cable, corroded contact pins on the steri-mate handpiece.Cabinet has broken body posts.Will replace damaged/worn components and recalibrate unit to factory specs upon estimate approval.Dhr review is not required because the product was returned for evaluation, and the described failure mode is a known hazard.
|
| |
|
Search Alerts/Recalls
|