• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. BURR, ROUND, 8 FLUTE 4.0MM X 13CM; BURR,SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. BURR, ROUND, 8 FLUTE 4.0MM X 13CM; BURR,SURGICAL Back to Search Results
Model Number BURR, ROUND, 8 FLUTE 4.0MM X 13CM
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/18/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
On 30th december 2025, it was reported by an arthrex employee via email that for 2 units of ar-8400rbe, round burr, 8 flute, 4.0 mm x 13 cm, the spherical grinding head were shedding a large amount of chips.This was detected during use in an ankle joint repair surgery on (b)(6) 2025.The procedure was completed successfully as another brand's product was used.There was no patient harm and no secondary procedure needed.No dualwave outflow tubing used in the case, there was an uninterrupted flow of irrigation through the device during its use.Shaver consoles and handpieces type and serial number is (b)(6).
 
Additional Manufacturer Narrative
Additional information: d9, g3, h3, h6.Complaint allegation is confirmed.Visual inspection noted chipping at the edges of the inner tube tip.Chipping was also observed along the edges of the slots and at the tip of the outer tube.Per dfu-0215-eo- f precautions 6.Do not allow the rotating portion of any device to touch any metallic object, such as a cannula or arthroscope, during use.Damage to both devices is likely.Damage to the device can range from a slight distortion or dulling of the blade/burr edge to actual fracture of the tip in vivo.If such contact does occur, inspect the tip.If there are cracks, fractures or dulling, or if there is any other reason to suspect a blade is damaged, replace it immediately.7.Excessive "side-loading" on the device during use does not improve cutting performance and in extreme cases will cause irreversible damage to the device.The most likely cause is attributed to misuse of the device, including exposure to excessive mechanical forces, misaligned insertion, side loading during use and/or allowing the rotating portion to touch any metallic object during use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BURR, ROUND, 8 FLUTE 4.0MM X 13CM
Common Device Name
BURR,SURGICAL
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key24067557
Report Number1220246-2026-00202
Device Sequence Number10282766
Product Code GFF
UDI-Device Identifier00888867043626
UDI-Public(01)00888867043626(11)240812(17)290731(10)15300568
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 04/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBURR, ROUND, 8 FLUTE 4.0MM X 13CM
Device Catalogue NumberAR-8400RBE
Device Lot Number15300568
Was Device Available for Evaluation? Yes
Distributor Facility Aware Date12/30/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/30/2025
Supplement Date Received by Manufacturer04/28/2026
Initial Report FDA Received Date01/14/2026
Supplement Report FDA Received Date04/30/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/12/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-