| Catalog Number |
PN-0003000 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
| Date of Event |
12/15/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Tablo instructions for use (ifu) warning and caution: when the blood in dialysate alarm occurs, the dialysate must be tested for the presence of blood, and the system alarm screen prompt must be answered to clear the alarm.Follow the alarm screen instructions to sample the effluent dialysate and test the sample as instructed by your center.The system will prevent treatment continuation in the event of the presence of blood in the dialysate.Outset technical support engineer (tse) reviewed the system log with the procedure date (b)(6) 2025 and confirmed two instances alarm_dialyzer_blood_leak.No console-related malfunctions or performance issues were identified.Furthermore, an outset field service engineer (fse) conducted an onsite mock treatment and verified that the blood detector monitor operated within specification.A review of production records for this serial number did not note any related manufacturing nonconformances that would contribute to this event.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a blood line leak alarm during patient treatment.Blood strips did confirm that there was a leak.Blood was discarded twice during treatment, resulting in an estimate of 600 ml blood loss.No patient adverse event was noted as a result of this incident.
|
| |
|