| Model Number |
305U25 |
| Medical Device Problem Codes |
Calcified (1077); Patient Device Interaction Problem (4001); Intravalvular regurgitation (4068)
|
| Health Effect - Clinical Codes |
Calcium Deposits/Calcification (1758); Dyspnea (1816); Fatigue (1849); Heart Failure/Congestive Heart Failure (4446); Aortic Valve Insufficiency/ Regurgitation (4450); Swelling/ Edema (4577)
|
| Date of Event |
12/29/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that 16 years and 8 months post implant of this 25mm aortic bioprosthetic valve, the patient is being evaluated for a transcatheter valve-in-valve replacement.The reason for evaluation was reported per echocardiogram as moderate to severe aorticregurgitation.It was reported that patient also suffered decompensated heart failure, shortness of breath on exertion (sob), fatigue and lower extremity edema.Evidence of calcification was seen on the left ventricular outflow tract (lvot) under the non-coronary cusp (ncc) and in the proximal right coronary artery (rca), sino-tubular junction (stj) and ascending aorta.Subsequently on (b)(6) 2025 the patient successfully had an aortic valve-in-valve replacement with a non-medtronic transcatheter valve.No additional adverse patient effects were reported.
|
| |
|
Search Alerts/Recalls
|